G. Mendell, S. Karki, D. Tuyenbayev
Attached are plots showing the calibration factors (kappa_tst, kappa_pu, kappa_A, kappa_C and f_c) for 26 Sept. 2016 generated from the Spectral Line Monitor (SLM) data, analyzed using Matlab code from Sudarshan Karki and EPICS values from Darkhan Tuyenbayev found in
D20160926_H1_CAL_EPICS_VALUES.m
in the calibration svn under,
aligocalibration/trunk/Runs/PreER10/H1/Scripts/CAL_EPICS
The plots start at 1158913200 == Sep 26 2016 08:19:43 UTC
A full set of plots can be found by going here,
https://ldas-jobs.ligo-wa.caltech.edu/~gmendell/pcalmon_with_plots/daily-pcalmonNavigation.html
and clicking on the calendar in the left frame, and then any of the links to the plots in the middle frame. The plots for Sept 26, 27, and 28 have been updated using D20160926_H1_CAL_EPICS_VALUES.
(Plots and data have not shown up after Sept 28, due to a configuration problem introduced during maintenance on 27 Sept 2016. This is also now fixed and SLM should produce plots again, when we have more H1 analysis ready data.)
See also these related alogs:
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=29992
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=29184