H1 PSL
jason.oberling@LIGO.ORG - posted 09:55, Tuesday 18 July 2017 (37579)
PSL Weekly FAMIS Tasks (FAMIS 3659 & 8431)

This morning I performed the weekly PSL FAMIS tasks.

HPO Pump Diode Current Adjust (FAMIS 8431)

I adjusted the HPO pump diode currents, changes summarized in the table below.  All adjustments were done with the ISS OFF.  I have attached a screenshot of the PSL Beckhoff main screen for future reference.

  HPO Diode Currents (A)
Old New
DB1 48.9 49.0
DB2 51.8 52.0
DB3 51.8 52.0
DB4 51.8 52.0

I then tweaked the DB operating temperatures.  The only change was to DB1, all diodes changed from 29.0 °C to 28.5 °C; no changes were made to the operating temps of DBs 2, 3, or 4.  The HPO is now outputting 154.9 W and the ISS has been turned back ON.  This completes FAMIS 8431.

PSL Power Watchdog Reset (FAMIS 3659)

I reset both PSL power watchdogs at 16:26 UTC (9:26 PDT).  This completes FAMIS 3659.