H1 AOS (AOS, ISC, SUS)
keita.kawabe@LIGO.ORG - posted 13:32, Friday 11 May 2018 - last comment - 11:04, Monday 14 May 2018(41954)
TMSX SUS rotation complete (Corey, Travis, Keita)

Betsy and Travis told us that we can just use the top blade to rotate TMS is we're talking about O(100urad) instead of O(many mrad), so we did just that.

The procedure was: Remove ERM protection cover, adjust EX bias so the optic is centered in oplev, inject green laser, set the TMSX offset to get the retro-reflection on the table. Then loosen the blade (which makes TMS sag so no retro-reflection), turn the adjustment screw, tighten the blade, look where the beam goes, repeat.

It turns out that 1/4 turn of the adjustment screw took out 400-something urad, then Travis tried 1/16 of a turn and that took out the rest.

In the end TMSX alignment sliders were set to [P,Y]=[-74,5] urad, almost perfect for Y, and P is no worse than yesterday where it was [91,-585].

We also adjusted BOSEM depth. For F1, LF and RT we recently measured the open values and set the depth (41675) so didn't bother to remeasure open values but we reset the depth anyway as RT seemed to have drifted out.

For F2, F3 and SD, old and new offset as well as open value as of now are:

  Open today New offset Old offset Reading before adjustment Reading after adjustment
LF NA NA -9350 ~9.9k ~9k
RT NA NA -13475 ~16k ~12k
F1 NA NA -10515 ~10.9k ~10.7k
F2 18.1k -9050 -11164 ~11k ~9k
F3 23.2k -11600 -14150 ~18.5k ~11.3k
SD 22.1k -11050 -14372 ~11k ~11.5k

 

Inventory info: S/N of F3 BOSEM is 296, F2 301.

Comments related to this report
betsy.weaver@LIGO.ORG - 14:09, Friday 11 May 2018 (41957)

Travis and I have taken the 6 DOF transfer function measurements of TMSX.  While nothing says it is rubbing (good), there still seems to be something wrong with the top 2 BOSEMs (LF and RT), see attached.  The TFs that utilize these 2 BOSEMs show a DC offset across the entire measurment.

Will continue forensics later.  Maybe a setting somewhere?  TBC...

Images attached to this comment
Non-image files attached to this comment
keita.kawabe@LIGO.ORG - 11:04, Monday 14 May 2018 (41972)

The changes in the TF look predominantly just a factor of 4-ish increase in sensitivity, and it makes sense to me.

We know that BOSEM LF and RT sensitivity was much smaller before I and Corey repositioned/replaced these on Apr/25. If you look at my alog and Jeff's comment attached (41675),  their physical sensitivity increased by a factor of 2.8 and 3.3 respectively for LF and RT, then Jeff changed the calibration constant for LF from 1.295 to 1.604 (a factor of 1.24) and for RT from 1.146 to 1.113 (a factor of 0.97).

The overall sensitivity of LF and RT increased by 2.8*1.24=3.5 and .97*3.3=3.2, respectively, i.e. roughly the P and V sensitivity should have increased by a factor of ~(3.2+3.5)/2~3.4. And 4/3.4~1.2, that's not a big number.

I assume that the reference in Betsy's plot was taken when the sensitivity was already small.