[Stefan, Lisa, Kiwamu, Yuta, Evan]
We are trying to center the transmitted X-arm IR on the transmon QPDs, so that we can have an IR alignment reference. The expected power on each QPD is determined as follows:
These numbers give an expected power of 3.7 µW on each QPD.
What we measured was about 600 cts for the sum output on each QPD. We can back out the power on the QPDs as follows:
This gives a power of 4.6 µW on each QPD.
[Keita, Yuta, Evan]
Some corrections:
So the expected and measured numbers are actually much closer.