Transmon X QPD power budget

[Stefan, Lisa, Kiwamu, Yuta, Evan]

We are trying to center the transmitted X-arm IR on the transmon QPDs, so that we can have an IR alignment reference. The expected power on each QPD is determined as follows:

These numbers give an expected power of 3.7 µW on each QPD.

What we measured was about 600 cts for the sum output on each QPD. We can back out the power on the QPDs as follows:

This gives a power of 4.6 µW on each QPD.

Comments related to this report
evan.hall@LIGO.ORG - 10:22, Friday 14 February 2014 (10099)

[Keita, Yuta, Evan]

Some corrections:

  • The ETM transmissivity for red is 3.6 ppm (from the nebula page). The transmon table has four silver-coated mirrors, each with reflectance between 97% and 98%. Including these factors in the calculation gives an expectation of 2.4 µW.
  • In the measured count calculation, I forgot to include the 45 dB whitening gain and somehow goofed on the final number. Putting in this gain gives a measurement of 2.6 µW.

So the expected and measured numbers are actually much closer.