Reports until 21:44, Wednesday 12 February 2014
H1 SUS (IOO, ISC)
jeffrey.kissel@LIGO.ORG - posted 21:44, Wednesday 12 February 2014 (10043)
H1 SUS MC2 M2 COILOUTF
J. Kissel

I've found a window to stick in the proper coil driver compensation filters into MC2. As described in LHO aLOG 9956, the compensation now correctly compensates the modified M2 triple acquisition driver,

FM1    zpk([13],[65],1,"n")
FM2    zpk([11;21],[1;210],1,"n")
FM6    zpk([65],[13],1,"n")
FM7    zpk([1;210],[11;21],1,"n")
The Binary IO filter logic had been changed long ago, so everything works as designed now. Also, I've cleaned up and removed what was in FMs 3 and 8, which was redundant with FM6 and FM7. The new filter file has been committed to the userapps repo.