Reports until 13:00, Friday 14 February 2014
H1 CDS (SEI, SUS)
david.barker@LIGO.ORG - posted 13:00, Friday 14 February 2014 (10106)
new h1iopsusey model, removed DC component from the Software Watchdog
WP#4447
Created new h1iopsusey model to remove the DC components of the ETMY software watchdog. At the same time I did the same for TMSY though this is not used in the WD trigger circuit. Related changes to the safe.snap and the MEDM screen were made. at 12:17 local time the h1susey models were stopped to install the new IOP model. I tested the WD using the PANIC button on the SUS DACKILL. I then restarted the h1susetmy and h1sustmsy models, burt restoring to 11am today.
A DAQ restart is needed for the IOP model, can wait till later.