daniel.sigg@LIGO.ORG - posted 16:36, Wednesday 19 February 2014 - last comment - 14:33, Thursday 20 February 2014(10190)
RT comm link in EX
After re cabling for the ALS WFS the link between slow and fast controls stopped working. The newly assigned DAQ ADC channel has a large -5V offset and seems broken. The offset is there even if nothing is connected to the AA chassis.
Comments related to this report
daniel.sigg@LIGO.ORG - 14:33, Thursday 20 February 2014 (10218)
Changing the AA chassis didn't fix the problem. So, it is probably the ADC. To minimize the disruption we simple switched to a different channel for now.