Laser Status: SysStat is good Output power is 28.9 W (should be around 30 W) FRONTEND WATCH is Active HPO WATCH is red PMC: It has been locked 1 day, 22 hr 20 minutes (should be days/weeks) Reflected power is 1.1 Watts and PowerSum = 11.9 Watts. (Reflected Power should be <= 10% of PowerSum) FSS: It has been locked for 0 d 4 h and 5 min (should be days/weeks) Threshold on transmitted photo-detector PD = 0.89 V (should be 0.9V) ISS: The diffracted power is around 6.9 % (should be 5-15%) Last saturation event was 0 d, 4 h and 30 minutes ago (should be days/weeks)