Displaying report 1-1 of 1.
Reports until 19:44, Monday 24 February 2014
H1 ISC
keita.kawabe@LIGO.ORG - posted 19:44, Monday 24 February 2014 - last comment - 19:52, Monday 24 February 2014(10293)
Green WFS Monday (Jax, Stefan, Keita)

After painful trial and error process we've found a solution that converges.

Comments related to this report
stefan.ballmer@LIGO.ORG - 19:52, Monday 24 February 2014 (10294)
We hooked up H1:ALS-X_WFS_DOF_2_P_OUTPUT to the end X PDH servo board, EXC A input. This allowed us to drive a clean length signal, providing a nice way to set all demod phases.

In the end the servos converged nicely, but we noticed that the power went up some with a -200ct offset in DOF_2_P. RFAM?

We also tried guessing the output matrix by sticking in an error offset. However the underlying drift proved too big for a clean result.

Displaying report 1-1 of 1.