(Alexa, Jax, Stefan)
Alexa was trying to hunt down a missing gain factor for her model and instead found a busted COMM PLL board (serial number S1200564). The op-amp for the VCO compensation stage (U37) was blown. I recalled replacing the same op-amp on the DIFF PLL about a month ago, suggesting that there's something that makes this particular op-amp vulnerable. After discussing with Richard McCarthy, we opted to change resistor R96 from 0 to 2k ohms to protect the poor thing. Lab tests show that it's in good shape, and it'll be back to the ISC rack as soon as we're done with the ISC meeting.
Should this resistor change be propagated to all PLL boards?
No, we're still thinking about it. It's going in as modified and if it keeps having problems, we'll burn that bridge when we get to it.