Reports until 16:29, Friday 28 February 2014
H1 ISC
daniel.sigg@LIGO.ORG - posted 16:29, Friday 28 February 2014 (10430)
79.2 MHz intermodulation product

With the RF doubler for the 79.2 MHz source moved on top of the ALS distribution chassis, we took another look at the modecleaner error signal. The intermodulation product with the PSL VCO is still visible but small. It is about an order of magnitude smaller than before. This seems good enough for now, but could potential effect the noise of the full interferometer.