Displaying report 1-1 of 1.
Reports until 09:11, Tuesday 04 March 2014
H1 General
andres.ramirez@LIGO.ORG - posted 09:11, Tuesday 04 March 2014 (10485)
PSL Check
PSL Check:
Laser Status: 
SysStat is good
Output power is 28.7 W (should be around 30 W)
FRONTEND WATCH is Active
HPO WATCH is red

PMC:
It has been locked 2 days, 15 hr 39 minutes (should be days/weeks)
Reflected power is 1.2 Watts  and PowerSum = 11.9 Watts.
(Reflected Power should be <= 10% of PowerSum)

FSS:
It has been locked for 0 d 9 h and 51 min (should be days/weeks)
Threshold on transmitted photo-detector PD = 0.55 V (should be 0.9V)

ISS:
The diffracted power is around 1.5 % (should be 5-15%)
Last saturation event was 0 d, 10 h and 1 minutes ago (should be days/weeks)