Displaying report 1-1 of 1.
Reports until 00:29, Wednesday 05 March 2014
H1 ISC
sheila.dwyer@LIGO.ORG - posted 00:29, Wednesday 05 March 2014 - last comment - 13:38, Wednesday 05 March 2014(10511)
measurement of IR noise with COMM locked

Alexa, Daniel, Sheila

We normalized the refl PDH error signal, and saw that we can stay in the linear range of this signal with only ALS COMM locked. 

We first normalized LSC-X_TR_A_LF by adding a gain of 0.000154 so that the tranmission peak is at roughly 1.  We set LSC-REFLBIAS to  REFL_A_RF9_I over the transmitted power (LSC-X_TR_A_LF).  The first screen shot is the PDH, the transmitted power, and the normalized error signal as we move the COMM VCO set frequency.  In the red trace we moved the frequency by 2kHz; meanwhile, in the blue trace we moved half as fast over 1kHz.  We used Alexa's equation from alog 7054  to roughly calibrate the signal: 80Hz cavity pole/150 counts pp of the PDH signal=0.533Hz/count.

We put this gain into the REFL_BIAS filter, we will do a more carefull calculation in the morning. With this calibrated error signal we were able to measure an out of loop spectrum of the noise between the arm cavity and the PSL.  Attached are some of the spectra that we took. In the top panel the green trace is a reference with the cleanroom over HAM1 on, in the red trace we turned the cleanroom off and Alexa tried to tune up the arm cavity alignment by maximizing the green transmitted power.  This realingment clearly helped at low frequecies, and reduced the coherence with the ETMX PIT oplev. 

The RMS we measured is completely dominated by low frequency noise that changed, probably because the alignment was drifting.  The RMS down to 0.1 Hz varied from 30Hz-70Hz, we also measured an RMS of 58Hz down to 0.01 Hz.

We saw coherence with the end station PDH reflected signal above 1kHz,  coherence with the PIT oplevs (mostly ETMX, which currently has level 2 controlers on Stage 1 and damping only on stage 2) around the PIT resonances (0.4-0.5 Hz), and coherence with YAW OpLevs from 0.04-0.5 Hz.  We also looked for coherence with some of Robert's PEM sensors, we have coherence from the periscope on ISCT1 from about 50-100Hz, as we kind of expected.

Tonight the noise is much worse than it was on saturday, the transmitted power was not stable at all over time.  This might be because we don't have as good of an alignment, or be because the arm cavity motion is larger because of the problems with ETMX ISI.  We also measured the spectrum with higher and lower gains in the COMM board, higher gain did not seem to help, which is different from the situation on saturday alog 10439

Images attached to this report
Comments related to this report
peter.fritschel@LIGO.ORG - 06:44, Wednesday 05 March 2014 (10516)

Did you have the ALS WFS servos running for these measurements?

jeffrey.kissel@LIGO.ORG - 09:07, Wednesday 05 March 2014 (10520)
Note, this same measurement at LLO looks strikingly similar: see LLO aLOG 11265!
sheila.dwyer@LIGO.ORG - 13:15, Wednesday 05 March 2014 (10551)

No WFS, dither alignment, or longitudnal feedback to the test mass was on for any of these measurements. 

daniel.sigg@LIGO.ORG - 13:38, Wednesday 05 March 2014 (10554)

When comparing this spectrum to the LLO one has to undo the red cavity pole in the above spectra. It is a little bit more complicated, since we are dividing by the red transmitted power which itself can carry noise. With this correction both spectra are referenced to the input laser frequency and look more or less identical.

When comparing this spectrum to the HIFO-Y spectrum in T1300688, the story at mid frequencies is more or less the same:

  • There is large periscope peak around 70Hz with no evidence that the crossfire beam propagation solution helped at all.
  • Between 100Hz and 1kHz we are limited by acoustic noise from the corner.
  • At frequencies of a 1kHz and above we see the laser noise from the end station. Due to the much lower green finesse, it shows up an order of magnitude stronger.
  • At frequencies below 10Hz we now see a strong angular coupling, again due to the lower green finesse. The wavefront sensors were not engaged during the above measurement, and the excess is about two orders of magnitude over the HIFO-Y baseline.
Displaying report 1-1 of 1.