Displaying report 1-1 of 1.
Reports until 16:19, Wednesday 05 March 2014
H1 INS (SEI)
jim.warner@LIGO.ORG - posted 16:19, Wednesday 05 March 2014 (10561)
ETMY ISI rebalanced, measurments starting tonight

It was realized yesterday that TMS still had some "head-banger" hardware left from the cartridge install. This was removed today, which necessitated a rebalancing of St2. While working on this, I also looked at a CPS that was acting like it was broken. I poked around the cable for the CPS and it fixed itself, so I'm hoping it was just a loose connection. I'll look at this again tomorrow, but at this point it looks like the in-air cable MAY be bad. For now, I will be running TF's tonight. from OPSWS5.