Sheila, Jeff, Jim
We have added OpLev BLRMS to the suspension models for suspensions that have opLevs, this change is described in ECR E1400155
This involved changes to h1susetmx h1susitmx (since these are not using the quad master right now) QUAD_MASTER BSFM_MASTER and HLTS_MASTER
There is also a new medm screen in sus/common/medm/SUS_CUST_OPLEV_BLRMS
I also added a link to this screen in the op Lev screens :SUS_CUST_QUAD_L3_OPLEV.adl SUS_CUST_HLTS_M3_OPLEV.adl and SUS_CUST_BSFM_M3_OPLEV.adl
All of these changes are now committed to the svn.
The models are running, and Jim has restarted the DAQ so the new channels are avaiable. We also checked using dtt and a spectrum that the values diplayed are accurate.
I also placed two strip tools, the top one for at the microseism frequency, 30-100mHz, and one for the suspension resonance frequencies, 0.3-1Hz. on the monitor on the wall next to the PEM FOMs. I moved the Osem screen over to the other side of th control room. We can change around the settings on these plots to make them more usefull if it turns out they are too hard to read.