Displaying report 1-1 of 1.
Reports until 19:01, Wednesday 19 March 2014
H1 AOS (SUS)
jeffrey.kissel@LIGO.ORG - posted 19:01, Wednesday 19 March 2014 - last comment - 13:48, Tuesday 25 March 2014(10881)
H1 ETMX Optical Lever Confusion
Has the analog whitening settings changed on the ETMX optical lever? I've compared a high-frequency spectrum with a measurement I took in February, and there's reported tons more motion above 2 [Hz], when the ISI motion is roughly equivalent if not better at these frequencies (see attached). Looks there's an extra 1:10 analog filter that's not being compensated or something...
Non-image files attached to this report
Comments related to this report
arnaud.pele@LIGO.ORG - 15:28, Thursday 20 March 2014 (10904)

Looking at a previous OPLEV spectra that was made in February on ETMX ITMX and ITMY (calibrated in urad/sqrtHz), it looks like the signal amplitude @ 10Hz (on the three of them) was similar to what you have on your red curve (~0.1 nrad/sqrtHz). Maybe the green trace (reference) was taken with a digital compensation filter engaged, but no analog whitenning ?

thomas.vo@LIGO.ORG - 13:48, Tuesday 25 March 2014 (10988)
Refer to ALOG-10267 for the changes in whitening settings on Feb 21st 2014. This accounts for the two orders of magnitude difference you see above 10 Hz.