H1 ISC
alexan.staley@LIGO.ORG - posted 12:19, Tuesday 01 April 2014 (11109)
Comm handoff measurements

We have some new settings for the Comm Handoff, so I repeated the usual measurements:

COMM PLL Servo Board

CM Servo Board

LSC-CARM Filter

LSC-REFL-SERVO-SLOW

MC2_M3_LOCK_L

MC2_M2_LOCK_L

MC2_M1_LOCK_L

MC2_M3_ISCINF

  1. Common path OLTF: Handoff_COMM_OLTF_mag/phase.txt
  2. Inject into fast path, cf/1-cs OLTF: Handoff_cfast_OLTF_map/phase.txt
  3. Inject into slow path, cs/1-cf OLTF: Handoff_cslow_OLTF_mag/phase.txt
  4. Crossover using LSC_CARM inputs: /ligo/home/sheila.dwyer/ALS/HIFOX/COMM/Crossover_April1.xml

The UGF of the comm handoff is now 3kHz with 80 deg phase margin. The crossover between the fast and slow path is at 13 Hz with 19 deg phase margin.