End Y PLL Open Loop TF results:
Servo Board Settings:
The UGF is at 23kHz with a phase margin of 52 deg (consistent with HIFOX). Using the AG spectrum analyzer, I tried finding the PZT resonance. The data is a bit ratty, but there appears to be a resonance a little below 200kHz (see the picture -- the two pictures were collected with different BWs; the first at 300Hz and the second at 2Hz; this notch-looking feature seemed to be repeatable...). Meanwhile, the board has a notch at around 280kHz, which also seems to appear in the pictures.
I have also uploaded and plotted the data I collected.