I noticed the frontend watchdog was still in FAIL state from last restart---I went into the H1Diode room and reset watchdog.
Laser Status: SysStat is good Output power is 27.9 W (should be around 30 W) FRONTEND WATCH is Active HPO WATCH is red PMC: It has been locked 3 days Reflected power is 1.4 Watts and PowerSum = 11.7 Watts. (Reflected Power should be <= 10% of PowerSum) FSS: It has been locked for 12 min (should be days/weeks) Threshold on transmitted photo-detector PD = 0.82 V (should be 0.9V) ISS: The diffracted power is around 2.54 % (should be 5-15%) Last saturation event was 13 minutes ago (should be days/weeks)