H1 SEI (SEI)
richard.mittleman@LIGO.ORG - posted 19:34, Friday 11 April 2014 - last comment - 08:25, Monday 14 April 2014(11307)
ETMY ISI Needs some TLC

ETMY isn't working as well as ETMX.

ETMY is running TCrappy blends

ETMX is running TBetter with sensor correction (lower blend frequencies in rX and rY)

  There is what looks like a loop instablity around 10Hz (stage 1 maybe?)

  To much motion in ry & rX around 1/2 (probably the difference in the blend frequencies)

 

  The references in the ETMX-GS13 and ETMX OPLev plots are with Sensor correction off, which is giving more peaking at the microsiesem then we want so I'm turning it off

 

Images attached to this report
Comments related to this report
richard.mittleman@LIGO.ORG - 08:25, Monday 14 April 2014 (11314)

The low frequency excess motion is likely due to the fact that we haven't finished doing the stage 1 tilt decoupling, Jim needs a few hours to finish that up