Laser Status: SysStat is good Output power is 27.5 W (should be around 30 W) FRONTEND WATCH is Active HPO WATCH is red PMC: It has been locked d, 11 h 40 minutes (should be days/weeks) Reflected power is 1.2 Watts and PowerSum = 11.0 Watts. (Reflected Power should be <= 10% of PowerSum) FSS: It has been locked for 0 d 0 h and 3 min (should be days/weeks) Threshold on transmitted photo-detector PD = 0.56 V (should be 0.9V) ISS: The diffracted power is around 12.4 % (should be 5-15%) Last saturation event was 0 d, 0 h and 2 minutes ago (should be days/weeks)