Displaying report 1-1 of 1.
Reports until 09:55, Sunday 20 April 2014
H1 AOS (DAQ)
david.barker@LIGO.ORG - posted 09:55, Sunday 20 April 2014 - last comment - 10:01, Sunday 20 April 2014(11461)
CDS model and DAQ restart report, Saturday 19th April 2014

model restarts logged for Sat 19/Apr/2014
2014_04_19 13:26 h1hpietmx
2014_04_19 13:27 h1hpietmx
2014_04_19 13:30 h1susetmx
2014_04_19 15:01 h1susetmy
2014_04_19 15:03 h1hpietmy
2014_04_19 15:18 h1susitmy
2014_04_19 15:19 h1hpiitmy
2014_04_19 15:29 h1susitmx
2014_04_19 15:32 h1hpiitmx
2014_04_19 15:45 h1sustmsx
2014_04_19 16:10 h1sustmsy
2014_04_19 17:14 h1hpibs
2014_04_19 17:19 h1susbs

2014_04_19 18:35 h1broadcast0
2014_04_19 18:35 h1dc0
2014_04_19 18:35 h1fw0
2014_04_19 18:35 h1fw1
2014_04_19 18:35 h1nds0
2014_04_19 18:35 h1nds1

A busy commissioning day followed by a DAQ restart. No unexpected restarts.

Comments related to this report
david.barker@LIGO.ORG - 10:01, Sunday 20 April 2014 (11462)DAQ, SEI

I just cleared the warnings raised by yesterday's restarts and all cleared except a DAQ error on HEPI BS and ETMY. Looks like their INI files are out of sync between FE and DAQ (model restart post-DAQ-restart). We might want to restart the DAQ ASAP as data from these systems is most probably corrupt.