(Keita, Alexa)
A more in depth alog will be posted later about what we did today at EY. For now I have attached the data for the OLTF we measured (mag --> 122.txt, phase --> 123.txt). The UGF was about 2.3kHz with a phase margin of a little less than 60 deg. The phase shifter was adjusted to 23ns. The PDH REFL servo board settings were as follows:
We also measured the peak to peak of the PDH error signal and found it to be ~816mVpp.
We remeasured the green power along ISCTEY (its possible we had the wrong laser setting on the powermeter when we measured this last):