A pressure sensor at MY had been MAJOR Alarming the last few days, but this is due to it being worked on. It will be displaying values of a higher pressure for a month or so while it is valved out and being worked on. To prevent getting alarms from this, we've bumped up the MAJOR alarm setting to 1x10^-6 (originally it would alarm at 5.11x10-8).
Since the Vacuum computer was really slow, Dave ssh-ed into the computer (still was slow) & did the following:
vacuum@control0:~$ caget HVE-MY:Y5_246BTORR.HIHI
HVE-MY:Y5_246BTORR.HIHI 5.11e-08
vacuum@control0:~$ caput HVE-MY:Y5_246BTORR.HIHI 1.00e-06
Old : HVE-MY:Y5_246BTORR.HIHI 5.11e-08
New : HVE-MY:Y5_246BTORR.HIHI 1e-06
We must revert the alarm setting when the VAC work is complete.
OK, 1e-6 wasn't good enough (Kyle started baking out RGA components today & this shot up the pressure on this sensor). John said I could up the Alarm setting to 1e-05, and if it alarms here, then the VAC crew will just power OFF the sensor.