Reports until 09:47, Monday 28 April 2014
H1 PSL
patrick.thomas@LIGO.ORG - posted 09:47, Monday 28 April 2014 (11593)
PSL status check
Laser is on
Output power is 27.4 W (should be about 30 W)
Watchdog is red (not active?) (should be active)
No warnings in PSL SYSSTAT.adl except "VB program online"

PMC
Last relock 6 days ago (should be days/weeks)
Reflected power is 14.4% of transmitted power (should be 10% or less)

FSS
Locked for 2 days, 4 hours (should be days/weeks)
Trans PD threshold is .4 V (should be at least .9 V)

ISS
Diffracted power is ~ 9.6% (should be around 10%)
Last saturation event was 2 days, 5 hours ago (should be days/weeks)