PFD Cleared

(Sheila, Alexa)

We had thought that maybe the 1/f noise seen in COMM came from the PFD. Today we have ruled that out. We did this by replacing the PFD with the spare IQ demod channels of the IMC. In this confirgaution we measured the beatnote signal to be 250mVpp, and set the COMM PLL Input gain to 31dB. This resulted in a loop with -18dB less gain than the nominal configuration. We then used the control signal to measure the noise.

I have attached a DTT with the following traces:

Clearly, the PFD does not seem to be the culprate. These measurements also suggest that we should lower the UGF of the COMM PLL loop.

With the demod in place and the input gain at 31dB, I also measured the amplitude spectrum out of the IMON of the demod. I have attached the data.

Images attached to this report
Non-image files attached to this report
Comments related to this report
daniel.sigg@LIGO.ORG - 17:56, Wednesday 30 April 2014 (11661)

The VCO has a visible non-linear response, see second page here. If there is significant rms in the control signal at higher frequencies, this could potentially explain the excess noise.