Displaying report 1-1 of 1.
Reports until 16:09, Tuesday 09 August 2011
X1 SUS
betsy.weaver@LIGO.ORG - posted 16:09, Tuesday 09 August 2011 - last comment - 16:19, Tuesday 09 August 2011(1182)
ITMy M0 BOSEM death
For whatever reason (forensics yet to take place) the F2 M0 BOSEM died around the power outage event last week (would be great if someone wanted to trend and append).  I'm swapping s/n 638 with 261.  261 was plugged in and reading OLV H2:SUS-ITMY_M0_OSEMINF_F2_OUTMON (in the dim room) of -6337, with H2:SUS-ITMY_M0_OSEMINF_F2_INMON reading 22899.  The dark light OLV (occulted by flag) is 31879.  OSEM input filters all enabled (nominal).  Set to 50% at 4:10 pt.
Comments related to this report
betsy.weaver@LIGO.ORG - 16:19, Tuesday 09 August 2011 (1184)
Also ran the yaw M0 diagonalization and saw 20dB separation between appropriate signals.
Displaying report 1-1 of 1.