D. Hosken, G. Grabeel, T. Vo We were able to fix the ring heater chassis EtherCAT wiring by referencing the drawings. We initially turned both ITM ring heaters to 30 Watts at 6:46 PM PT but it seems as though the maximum power that the drivers can put out is ~15-16 Watts so we changed the script that Greg wrote to run the power at 13.5 Watts during the high power portion and .5 Watts in the low power portion. I'll come in tomorrow morning to run some trends and take some measurements. John verified in the first hour that there were no spikes in pressure.