In light of the upcoming vent, it seemed like a good time test the ability of the test mass optical levers to recover from the venting, swapping and re-pumpdown. I've recentered the optical levers on 2014-06-02 17:05 UTC with the following status:
SUS values:
ITMX_Pit = 38.7
ITMX_Yaw = -48.9
ITMY_Pit = 193.4
ITMY_Yaw = -181.1
HEPI Isolated and biases are:
ITMX_RX = -6020
ITMX_RY = -45350
ITMX_RZ = 71250
ITMY_RX = 44920
ITMY_RY = 48020
ITMY_RZ = -23500
ISI Stg1 Isolated and biases are:
ITMX_RX = -17600
ITMX_RY = 27450
ITMX_RZ = 30450
ITMY_RX = 9760
ITMY_RY = -28600
ITMY_RZ = 6760
ISI Stg2 Isolated and biases are:
ITMX_RX = -4565
ITMX_RY = -680
ITMX_RZ = 7744
ITMY_RX = 19623
ITMY_RY = -12484
ITMY_RZ = 7496