Reports until 16:54, Wednesday 04 June 2014
H1 AOS
thomas.vo@LIGO.ORG - posted 16:54, Wednesday 04 June 2014 (12216)
ITMY Calibration

Jeff Kissel, Thomas Vo

Jeff pointed out to me today in light of the RH test this past weekend that the ITMY optical lever gain is a factor of 2 different than the ITMX optical lever.  We're inclined to trust the ITMX calibration because Keita made a pretty accurate measurement using the baffle PDs and the entire 4 km arm.  I also remember specifically changing ITMX &ETMX gains to match Keita's measurement, but I don't remember doing ITMY/ETMY. 

Anyway, Y-arm optical levers have to be checked once we get an arm again.