PSL is down today. Laser Status: SysStat: “Frontend lid open” and warning “Service mode” are red Output power is 0 W (should be around 30 W) FRONTEND WATCH is Red HPO WATCH is red PMC: It has been locked 0 days, hr 0 minutes (should be days/weeks)0 Reflected power is 0 Watts and PowerSum = 0 Watts. (Reflected Power should be <= 10% of PowerSum) FSS: It has been locked for 0 d 0 h and 0 min (should be days/weeks) Threshold on transmitted photo-detector PD = 0 V (should be 0.9V) ISS: The diffracted power is around 0.6 % (should be 5-15%)