Reports until 21:52, Tuesday 10 June 2014
H1 ISC
keita.kawabe@LIGO.ORG - posted 21:52, Tuesday 10 June 2014 (12289)
Many HAM6 ISC ground loops detected

Baffled to find many ground loops, then I remembered that I was baffled by this same thing in all chambers.

Tested some (not all) of the HAM6 ISC cables for ground loop according to T1200131.

Each in-air cable was disconnected from the chassis while still connected to the chamber feed through (other in-air cables were still connected to both the chassis and feed through), and the continuity between the cable shell and the chassis ground was tested using DVM beeper (any continuity=bad).

  cable # good/bad
OMC DCPDs 307 bad.
OMC QPDs 404 bad
OM1/TT1 236 good
OM2/TT2 237 bad
OM3/TT3 238 bad
QPD sled 232 good
AS_C QPD 233 bad

FYI, for all of the above cables, there was a continuity between pin13 and shell (it's supposed to be like that).

WFS (both DC and RF), OMC PZTs, OMCS and SEI cables were not tested but I'll test at least WFS and OMC PZT.

Anyway, this is not acceptable for the chamber containing our main read out, and should be fixed in chamber.