H1 TCS (TCS)
greg.grabeel@LIGO.ORG - posted 18:30, Friday 13 June 2014 - last comment - 16:45, Wednesday 18 June 2014(12365)
ITMy Ring Heater Forensic
Thanks to the efforts of SUS I was able to get in and look at the ring heaters of ITMy today. There are no apparent problems with breakage of either the glass formers or the macor retaining pieces. However, the center macor piece (D1002545) looks to have wandered off center quite significantly. I didn't have any pictures from that angle while the quad structure was still in chamber. So, unfortunately, I am unable to determine when it wandered off, but it definitely warrants looking at ITMx closely before it leaves BSC 3 if possible. 

It should be noted that this is the older style of ring heater, but the macor element standoff is the same throughout all but the most recent version.
Images attached to this report
Comments related to this report
greg.grabeel@LIGO.ORG - 09:10, Monday 16 June 2014 (12370)TCS
As requested a picture with a few labels, showing how the macor has wandered off the center point where it should be.
Images attached to this comment
greg.grabeel@LIGO.ORG - 16:45, Wednesday 18 June 2014 (12414)TCS
The old ring heaters on ITMy have been removed from the old quad structure. While I was packaging the upper and lower ring heater I noticed quite a bit of particulate coming out. Looks like glass, certainly warrants a closer inspection.
Images attached to this comment