ITMx Position Alignment - The Story so Far + Historical Look at ITMx/ITMy Pilots
IAS: J. Oberling
SUS: B. Weaver, T. Sadecki
SEI: J. Warner
 
We started the position alignments for the new ITMx on Tuesday.  I immediately noticed the ITMx seemed to be sitting very high with respect to the ACB.  In fact, it was difficult to get a height measurement since it was very hard to make out the top edge of the optic.  After several rounds of adjustments over the next couple days (detailed by Betsy here and here, with Jim taking looks at HEPI here.) we are now out of vertical adjustment range on the quad (we started ~11.0 mm high and ended up 2.8 mm high after all was said and done).  The main issue I was having sighting the optic turned out to be a differential height between the ITMx and the CPx; I thought I was sighting the bottom edge of the ITMx, but it turned out to be the bottom edge of the CPx (we put something between the chains to isolate the ITMx edge yesterday; this was our 1st truly accurate measure of the ITMx height: +7.1 mm)  The remaining vertical error will need to be taken out with HEPI, which involves de-attaching the HEPI actuators, making the move, and re-attaching the HEPI actuators (as Jim noted, starting over with HEPI).  On a positive note the ITMx now looks to be much better centered vertically with respect to the ACB.  The current position errors of the ITMx are as follows:

I looked back through my notes of the pilot ITMx cartridge alignment from way back in September 2013 and the in-chamber alignment from November 2013 to see what the history of the height adjustments are with this quad.  The ITMx height on the test stand is set with respect to the ISI table: we attach a calibrated height stick (with attached scale) to the ISI table and use that to set the total station 1742.0 mm below the ISI table; this is the desired distance the center of the ITMx should be from the ISI table.  We then aligned the ITMx in height to this location.  We started low by 1.6 mm, but adjusted to be high by 0.1 mm.  The cartridge was then craned into WBSC3.  I seem to remember Hugh setting the ISI table to its proper height prior to the alignment work (correct me if I'm wrong here Hugh).  I then took the first set of numbers and found the ITMx to be 3.6 mm low.  We still do not understand where this discrepancy between the cartridge alignment, Hugh's setting of the ISI table, and my 1st in-chamber look at ITMx comes from, but I should note that for the pilot ITMx we needed to raise HEPI by 3.6 mm, now for the production ITMx we need to lower HEPI by 2.8 mm; Betsy floated the idea that it is a possibility that the pilot suspension could have been long and the production suspension is correct, thereby causing the required HEPI adjustments.

At Betsy's suggestion I also looked back at ITMy to see if we could be walking into a similar situation with this optic.  This suspension goes all the way back to WBSC8 in December of 2011.  When this suspension originally left the test stand we were 1 mm low.  After the in-chamber fiber break the ITMy was suspended by a wire for the H2OAT.  In November of 2012 this suspension was moved to WBSC1 to be used for HIFO-Y and the most recent integration work.  The ITMy on the test stand before install into WBSC1 had a height error of 0.2 mm low.  Once installed in WBSC1, we began aligning in January 2013; the initial height error was high by 1.7 mm.  We used HEPI to adjust the ITMy height error to 0.0 mm.  This was the last time we looked at ITMy (~18 months).  Looking at the other BSC chamber optics (BS, ETMx, ETMy), they all had initial height errors of <= 1.0 mm, and with the exception of the BS they all started high (BS was initially low by 0.4 mm, so no problems there).  The pilot ITMx in WBSC3 had the largest initial height error we've seen here at LHO, and was 1 of 2 optics to have a negative initial in-chamber height error.

Displaying report 1-1 of 1.