The MC1, 2 and 3 suspensions are now damped manually with the guardians paused for the HAM2 in-vac work. There was an issue with these guardians who reported some Ezca errors when they were in SAFE state. Arnaud and Mark will have a look at this issue.
For record purpose, here are the error messages in the SUS_MC2 guardian:
20140617_19:55:36.156 SUS_MC2 [SAFE]
20140617_19:57:31.994 SUS_MC2 W: if is_tripped(sustools.Sus(ezca)):
20140617_19:57:31.994 SUS_MC2 W: File "/opt/rtcds/userapps/release/sus/common/guardian/SUS.py", line 27, in is_tripped
20140617_19:57:31.994 SUS_MC2 W: trippedwds = susobj.trippedWds()
20140617_19:57:31.995 SUS_MC2 W: File "/opt/rtcds/userapps/release/sus/common/guardian/sustools.py", line 211, in trippedWds
20140617_19:57:31.995 SUS_MC2 W: result = []
20140617_19:57:31.995 SUS_MC2 W: File "/ligo/apps/linux-x86_64/cdsutils-238/lib/python2.7/site-packages/ezca/ezca.py", line 233, in read
20140617_19:57:31.995 SUS_MC2 W: pv = self._connect(channel)
20140617_19:57:31.995 SUS_MC2 W: File "/ligo/apps/linux-x86_64/cdsutils-238/lib/python2.7/site-packages/ezca/ezca.py", line 223, in _connect
20140617_19:57:31.995 SUS_MC2 W: pv.pvname))
20140617_19:57:31.995 SUS_MC2 W: EzcaError: Could not connect to channel (timeout=2s): H1:SUS-MC2_M1_WDMON_STATE
20140617_20:28:40.013 SUS_MC2 [SAFE]
20140617_20:28:42.076 SUS_MC2 [SAFE.run] ezca: connecting to ifo-rooted channel: H1:IOP-SUS_H34_DACKILL_STATE
20140617_20:28:46.083 SUS_MC2 W: if is_tripped(sustools.Sus(ezca)):
20140617_20:28:46.084 SUS_MC2 W: File "/opt/rtcds/userapps/release/sus/common/guardian/SUS.py", line 27, in is_tripped
20140617_20:28:46.084 SUS_MC2 W: trippedwds = susobj.trippedWds()
20140617_20:28:46.084 SUS_MC2 W: File "/opt/rtcds/userapps/release/sus/common/guardian/sustools.py", line 214, in trippedWds
20140617_20:28:46.084 SUS_MC2 W: trig = self.ezca.read(pv+'_STATE')
20140617_20:28:46.084 SUS_MC2 W: File "/ligo/apps/linux-x86_64/cdsutils-238/lib/python2.7/site-packages/ezca/ezca.py", line 233, in read
20140617_20:28:46.084 SUS_MC2 W: pv = self._connect(channel)
20140617_20:28:46.084 SUS_MC2 W: File "/ligo/apps/linux-x86_64/cdsutils-238/lib/python2.7/site-packages/ezca/ezca.py", line 223, in _connect
20140617_20:28:46.084 SUS_MC2 W: pv.pvname))
20140617_20:28:46.084 SUS_MC2 W: EzcaError: Could not connect to channel (timeout=2s): H1:IOP-SUS_H34_DACKILL_STATE
Arnaud's post from Friday implied that some of the watchdog channel names have changed. This is almost certainly what's going on here.
Fix the names in the guardian code! Presumably sustools.py needs to be updated.