Displaying report 1-1 of 1.
Reports until 12:53, Thursday 26 June 2014
X1 SUS
mark.barton@LIGO.ORG - posted 12:53, Thursday 26 June 2014 - last comment - 12:36, Tuesday 01 July 2014(12494)
QUAD08 TFs

Joe and Andres finished adjusting QUAD08 so I started a round of TFs on both chains at 12:52. Order is undamped M0, R0, then damped M0 and R0.

Comments related to this report
arnaud.pele@LIGO.ORG - 16:53, Thursday 26 June 2014 (12502)

First set of measurement failed. Started them again at 23:52:23 UTC.

arnaud.pele@LIGO.ORG - 15:46, Friday 27 June 2014 (12520)

Transfer functions ran yesterday in the staging building are showing clear sources of rubbing especially in the pitch/length/yaw degrees of freedom for the reaction chain. Main chain has large unexpected cross couplings in pitch.

arnaud.pele@LIGO.ORG - 12:36, Tuesday 01 July 2014 (12570)

Andres and Jeff realized the cover was on during that measurement, which is why results were bad. We started them again Friday overnight.