HAM4
In cleanroom start of session
All counts zero
In chamber start of session:
0.3um..10 counts
0.5um...10 counts
0.7um...10 counts
rest zero
In chamber end of session
(Didnt bother as was only in there for 2 minutes total)
HAM2
In cleanroom start of session
All counts zero
In chamber start of session
All counts zero
In chamber end of session
All counts zero
HAM3
In cleanroom start of session
0.3um...10 counts
rest zero
In chamber start of session
All counts zero
In chamber end of session
0.3um...350 counts
0.5um....180 counts
0.7um....110 counts
1.0um...70 counts
2.0um...20 counts
5.0um...0 counts
I took the readings 4 times and these are accurate. Mind you I guess it could be due to a combination of me moving baffles, and by this stage also work was going on in BSC2 trying to put the baffle on
I have included pics for people fo humidity/temp readings as well