Displaying report 1-1 of 1.
Reports until 15:20, Thursday 24 July 2014
H1 SEI
hugh.radkins@LIGO.ORG - posted 15:20, Thursday 24 July 2014 - last comment - 15:47, Thursday 24 July 2014(12971)
WHAM2 ISI Rebalanced. Left Locked for SUS Unlock

Not too bad to rebalance but only because everything needed was in the LVEA.  One of the four 10Kg Optical Table Masses(D0901075) had to be removed as there was so little wall mass on the West side of the ISI walls.  The bottom mass of the two stacker was removed and the upper damped mass was moved to the table top onto the damping viton.  So all three masses remaining on the table top are unbolted and resting on viton damping pads.

The total mass reduction of the Balance Mass Payload is 3.5kg.  Will update the drawing for this as-built.

Comments related to this report
hugh.radkins@LIGO.ORG - 15:47, Thursday 24 July 2014 (12973)

Have added an As-Built drawing to the D1000906-v3 DCC record.

Displaying report 1-1 of 1.