Displaying report 1-1 of 1.
Reports until 12:08, Friday 25 July 2014
H1 PSL
richard.mccarthy@LIGO.ORG - posted 12:08, Friday 25 July 2014 - last comment - 17:23, Friday 25 July 2014(12991)
ISS Photo Diode Array Test
Matt H.  requested that we verify the ISS PDs were still connected in vacuum.  I was able to get to the all 16 cables and tested each pair. They all tested fine with a reading of approximately .384VDC Anode to Cathode and OPEN Cathode to Anode on my Fluke 87V set to diode test. The last set I tested seemed to be reversed so I left them disconnected until I can have better access to the chassis under the chamber behind the HEPI support Structure.
Comments related to this report
matthew.heintze@LIGO.ORG - 17:23, Friday 25 July 2014 (13006)

Below is the instructions I received from Peter K and passed on. Thanks Richard for doing this

 

The only way to test all the diodes without laser light is to use the diode
tester function of a multimeter.  To do that you'll have to disconnect the
cables from the electronics that's where they were when we put the array
in.  With the multimeter on diode test, the negative (black) lead goes to the
cables/connector labelled cathode.  The positive (red) lead goes to the ones
labelled anode.  Touch the centre pins of the cable SMA connectors.  If
everything is okay you'll get a buzz and a voltage reading just over 0.3 V
(either 0.38 or 0.34, don't remember exactly which).

Displaying report 1-1 of 1.