H1 SUS
stuart.aston@LIGO.ORG - posted 12:59, Friday 01 August 2014 - last comment - 16:23, Tuesday 22 May 2018(13065)
ETMX (QUAD) L2 (PUM) UL channel excess sensor noise appears due to in-vacuum cabling
[Gary T, Danny S, Matt H, Stuart A, Norna R]

During the close-out work of the BSC9 chamber (see LHO aLOG entry 13085) we were able to further investigate the source of the excess noise that had previously been observed on the ETMX (QUAD) L2 (PUM) UL channel (see LHO aLOG entry 13047).

Previously, I had swapped field cables at the Satellite Box between ETMX L1 and L2 stages running to the chamber. This indicated that the noise was downstream, i.e. with the field cables themselves or in-chamber. We repeated this test, but this time swapping field cables at the air side of the vacuum flange. Further measurements then eliminated the field cables and confirmed the issue resided in-chamber.

Gary inspected and re-seated the in-chamber connectors from the vacuum flange to the OSEM, but this offered no improvement. Finally, the L2 UL AOSEM (s/n #270) was removed and a spare unit connected up in it's place, but again, unfortunately this offered no improvement.

This indicates that the excess noise is most likely a in-vacuum harness problem, which would be invasive to attempt to replace. Therefore, it was decided to generate an integration issue and proceed with the close-out of the chamber, with the excess sensor noise on this channel.
Images attached to this report
Comments related to this report
stuart.aston@LIGO.ORG - 13:00, Friday 01 August 2014 (13147)
Integration issue generated, see Bug #929.
travis.sadecki@LIGO.ORG - 16:23, Tuesday 22 May 2018 (42127)

Betsy and I swapped the quadrapuss cable to the L2 (PUM) AOSEMS.  This seems to have fixed the noise issue that had been plaguing the UL OSEM for several years.  See attached spectra for comparison.

After swapping the cable, we noticed that the 2 lower AOSEMs at the L2 stage (LL and LR) had low OLV values, so we decided to swap these as well.   We also took OLV values for UL and UR that were not replaced, but are dying of natural causes.

  Old S/N New S/N Old OLV New OLV Old Gain New Gain
LL 270 332 18726 18050 1.602 1.662
LR 483 473 25518 18150 1.176 1.653
UL N/A N/A 17934 21300 1.673 1.408
UR N/A N/A 25124 20600

1.194

1.456
Images attached to this comment