Displaying report 1-1 of 1.
Reports until 14:04, Tuesday 05 August 2014
H1 PSL (PSL)
peter.king@LIGO.ORG - posted 14:04, Tuesday 05 August 2014 (13216)
diagnostic breadboard scans
Diagnostic breadboard scans were executed; these are to measure the frequency noise, beam pointing,
mode content and relative power noise.

Some notes in passing:
 - The higher mode content of the frontend laser looks okay.
 - Autoalignment did not go green as expected, I will have to sort this out after reading the manual.
 - The beam pointing seems higher than the requirement but this could be because the calibration is out (plus my operating error).
 - The relative power noise is higher than the requirement, assuming the calibration is correct.
 
Non-image files attached to this report