Displaying report
1
-
1
of
1
.
Reports until 10:32, Monday 11 August 2014
H1 PSL (PSL)
Link
corey.gray@LIGO.ORG -
posted 10:32, Monday 11 August 2014
(13328)
Weekly PSL Check
Laser Status
Output power: 29.8W
FE Watchdog ACTIVE (HPO Watchdog still RED)
PSL SYS STAT: Looks FAIRLY normal ("VB Program Oline" & "
LRA Out Of Range
" are BOTH red)
PMC
Locked for last 1.5+ days
Reflected Power (5.8W) is NOT optimal and is higher than 10% of Total Power [25.5W] (Bad Alignment??)
FSS
Locked for last 1.5+ days
On video Refcav Refl looks like it may be off in yaw; Transmitted Power ("TPD [V]") = 1.14V (should be atleast 0.9V)
ISS
Locked; AOM Diffracted Power is around 9% (should be around 5-15%)
Last saturation event was 1.5+ days ago