Reports until 10:32, Monday 11 August 2014
corey.gray@LIGO.ORG -
posted 10:32, Monday 11 August 2014 (13328)
Laser Status
-
Output power: 29.8W
-
FE Watchdog ACTIVE (HPO Watchdog still RED)
-
PSL SYS STAT: Looks FAIRLY normal ("VB Program Oline" & "LRA Out Of Range" are BOTH red)
PMC
-
Locked for last 1.5+ days
-
Reflected Power (5.8W) is NOT optimal and is higher than 10% of Total Power [25.5W] (Bad Alignment??)
FSS
-
Locked for last 1.5+ days
-
On video Refcav Refl looks like it may be off in yaw; Transmitted Power ("TPD [V]") = 1.14V (should be atleast 0.9V)
ISS
-
Locked; AOM Diffracted Power is around 9% (should be around 5-15%)
-
Last saturation event was 1.5+ days ago