J. Kissel Completed a set of Phase 3a transfer functions and spectra to confirm the health of the OMCS before we close doors on Monday. Results look good, and comparable to previous measurements and LLO's OMC, sensors show noise comparable to what is expected. This SUS is clear for chamber close up. TF DTT Templates: /ligo/svncommon/SusSVN/sus/trunk/OMCS/H1/OMC/SAGM1/Data/ 2014-08-22_1739_H1SUSOMC_WhiteNoise_L.xml 2014-08-22_1739_H1SUSOMC_WhiteNoise_P.xml 2014-08-22_1739_H1SUSOMC_WhiteNoise_R.xml 2014-08-22_1739_H1SUSOMC_WhiteNoise_T.xml 2014-08-22_1739_H1SUSOMC_WhiteNoise_V.xml 2014-08-22_1739_H1SUSOMC_WhiteNoise_Y.xml TFs processed by /ligo/svncommon/SusSVN/sus/trunk/OMCS/Common/MatlabTools/plotOMCS_dtttfs.m TFs .mat file /ligo/svncommon/SusSVN/sus/trunk/OMCS/H1/OMC/SAGM1/Results/2014-08-22_1739_H1SUSOMC_M1.mat TFs compared with other previous results using /ligo/svncommon/SusSVN/sus/trunk/OMCS/Common/MatlabTools/plotallomcs_tfs.m Spectra taken by /ligo/svncommon/SusSVN/sus/trunk/OMCS/Common/MatlabTools/plotomcs_spectra.m ASD saved to /ligo/svncommon/SusSVN/sus/trunk/OMCS/Common/Data/2014-08-22_1739_H1SUSOMC_M1_Spectra.mat Spectra compared against others using /ligo/svncommon/SusSVN/sus/trunk/OMCS/Common/MatlabTools/plotallomcs_spectra.m All data and scripts committed to to SusSVN.