Displaying report 1-1 of 1.
Reports until 14:53, Tuesday 26 August 2014
H2 TCS
greg.grabeel@LIGO.ORG - posted 14:53, Tuesday 26 August 2014 - last comment - 10:27, Friday 29 August 2014(13602)
Removed Ring Heater and Cable Assemblies from H2 ITMy
Removed the Ring Heater Assembly from the ITMy quad structure going into 3IFO storage. Before starting I noticed the lower ring heater assembly had a crack in the glass former. Not sure when the break happened, but this one also displayed a lot of movement inside the ring heater shield. The pictures show how the former has pushed pretty far out of position. 

The ring heater cables were also removed. This includes the "antlers" (D1001755) but not the plate (D1002420) which the earthquake stop is attached to.

The assemblies removed were:

ASSY-D1001895-V5 #001 (lower ring heater, broken)
ASSY-D1001517-V7 #612 (cable assembly)
ASSY-D1001838-V6 #005 (upper ring heater)
Images attached to this report
Comments related to this report
greg.grabeel@LIGO.ORG - 19:20, Wednesday 27 August 2014 (13630)
That should read

ITMx (not y)

and

ASSY-D1001838-V6 #002 (upper ring heater)

not #005. typos.
betsy.weaver@LIGO.ORG - 11:11, Thursday 28 August 2014 (13645)

Note, when the 40kg optics are installed and removed from the QUAD lower structures that these ring heaters are mounted to, there is some torquing of the structure (and therefore the RH).  Possibly this adds to these failure modes.

greg.grabeel@LIGO.ORG - 10:27, Friday 29 August 2014 (13669)TCS
One more fix. The lower ring heater should be:

ASSY-D100195-108

Made a mistake with a V5 and V6 part.
Displaying report 1-1 of 1.