Reports until 17:50, Tuesday 26 August 2014
H1 AOS (SUS)
borja.sorazu@LIGO.ORG - posted 17:50, Tuesday 26 August 2014 - last comment - 07:24, Wednesday 27 August 2014(13612)
Checking ESD wiring at ETMX and comparing with ETMY

Borja

I just came back from a quick visit to End-X (UTC 2014-08-27 00:25:00).

I took some pictures (see attached) at the feedthrough wiring for the ESD at BSC9 (ETMX chamber). I compare it with the ones I took a while back at ETMY (see attached).

There is an obvious difference, the red insulating tape at ETMY wiring which does not show at ETMX. The wire which has the insulating tape carries VBIAS to the ESD at ETMY. The reason for the insulation is that the coax cable's shield is cut off because the corresponding coax cable inside the chamber has the known issue of having the shield connected to the center lead. This problem is non existen at ETMX.

This is another big difference between both masses which may be related to the more stable results on the charge measurements observed at ETMX.

Images attached to this report
Comments related to this report
rainer.weiss@LIGO.ORG - 07:24, Wednesday 27 August 2014 (13617)SUS
We need to fix the shielding in any event. The shield now acts as a large capacitance coupling of the electrode to the chamber and
will inevitably cause cross talk and noise in later operations.