Removed the Ring Heater Assembly from the ITMy quad structure going into 3IFO storage. Before starting I noticed the lower ring heater assembly had a crack in the glass former. Not sure when the break happened, but this one also displayed a lot of movement inside the ring heater shield. The pictures show how the former has pushed pretty far out of position. The ring heater cables were also removed. This includes the "antlers" (D1001755) but not the plate (D1002420) which the earthquake stop is attached to. The assemblies removed were: ASSY-D1001895-V5 #001 (lower ring heater, broken) ASSY-D1001517-V7 #612 (cable assembly) ASSY-D1001838-V6 #005 (upper ring heater)
That should read ITMx (not y) and ASSY-D1001838-V6 #002 (upper ring heater) not #005. typos.
Note, when the 40kg optics are installed and removed from the QUAD lower structures that these ring heaters are mounted to, there is some torquing of the structure (and therefore the RH). Possibly this adds to these failure modes.
One more fix. The lower ring heater should be: ASSY-D100195-108 Made a mistake with a V5 and V6 part.