Reports until 12:02, Monday 08 September 2014
H1 SEI
hugh.radkins@LIGO.ORG - posted 12:02, Monday 08 September 2014 (13810)
WHAM5 ISI TFs w/ HEPI Locked/At Air -vs- HEPI Isolating/Under Vacuum; and, compared to HAM4

Attached is a plot with the HAM5 CPS magnitudes from July 18 (at air HEPI locked) and from 6 Sept when we are under vacumm and the HEPI has closed position loops.  The things we believe are SUS resonances are cleaner now and the HEPI structure modes are lowered in frequency and Q. 

The feature just below 6hz is decidedly softer than the one seen on WHAM4--see comparison in the second attached plot.  All these 'HEPI' features are positioned/shaped differently between HAMs 4 & 5.  Maybe this is external mechanical and I should give everything a good hand-on look-see.

Images attached to this report