Following the recent replacement (generic maintenance) of PT120B, PT170B and PT180B Cold Cathode gauge sensor+electronics, there has been an apparent 25% discrepancy between the indicated pressures of the new PT170B and PT180B gauges -> Swapping the removable electronics between the two units has (surprisingly) resulted in both read backs being as expected in both absolute and relative pressure values